DT 2259A

Digital Stroboscope with fine tuning and measuring range upto 40,000 FPM

Features:
  • Automatic stop flashing to increase flashtube lifetime. 
  • Internal/external triggering and x2 / /2 button for fast adjustment. 
  • Uses photoelectric signal for external triggering, achieves automatic tracking function of the pattern. 


Applications:

This is used to provide image of single, two or multiple stillness for the vibrating object, rapidly rolling or a periodic motion object and can also observe movement tracks. Machine parts and vibrating string are common examples.


Measuring Principle:

This instrument consists of  a flashtube which produces brief repetitive flashes of light. Usually the rate of the stroboscope is adjustable to different frequencies. When a rotating or vibrating object is observed with the stroboscope at its vibration frequency (or a sub-multiple of it), it appears stationary. Thus, stroboscopes are also used to measure frequency. 

ModelDT 2259A
Display10mm 5-digits LCD
Measurement range50 ~ 40,000FPM
Accuracy±(0.05%n + 1d)
Resolution0.1FPM(50 ~ 999.9FPM), 1FPM(over 1000FPM)
Sampling time3 sec
External trigger level3 ~ 24V  with x2, /2 for fast check
H/L rangeFlashing light is much brighter at low range than at high range
Flash tube life100 million times
Regular operating conditionsNo corrosive gas and electromagnetic fields
Operating conditionsTemperature : 0 ~ 40°C, Humidity : <85% RH
Power supplyAC 220V
Power consumptionAbout 50W
Size and weight215 x 85 x 180mm, about 1kg
Standard accessoriesMain unit, manual, adapter, hard carry case
Catalogue: Catalogue